OTW BALLOON CATHETER & PKP BALLOON CATHETER
OTW balloon catheter includes three products: 0.014-OTW balloon, 0.018-OTW balloon, and 0.035-OTW balloon designed respectively for 0.014inch, 0.018inch, and 0.035inch guide wire. Each product consists of a balloon, tip, inner tube, development ring, outer tube, diffused stress tube, Y-shaped connector, and other components.
Excellent push
Complete specifications
Customizable
● Medical device products that can be used for processing include but are not limited to expansion balloons, drug balloons, stent conveyors, and other derivative products.
● Clinical applications include but are not limited to percutaneous transluminal angioplasty of the peripheral vascular system (including iliac artery, femoral artery, popliteal artery, sub knee artery, renal artery, etc.)
Unit | Typical Value | |||
0.014 OTW | 0.018 OTW | 0.035 OTW | ||
Guidewire Compatibility | inch | ≤0.0140 | ≤0.0180 | ≤0.0350 |
Sheath Compatibility | Fr | 4, 5 | 4, 5, 6 | 5, 6, 7 |
Usable Shaft Lengths | mm | 40, 90, 150, Customized | ||
Balloon Fold Configuration | 2, 3, 4, 5, 6, Customized | |||
Crossing Profile | mm | ≤1.2 | ≤1.7 | ≤2.2 |
Rated Burst Pressure (RBP) | atm | 14, 16 | 12, 14, 16 | 14, 18, 20, 24 |
Nominal Pressure(NP) | mm | 6 | 6 | 8, 10 |
Balloon Diameter | mm | 2.0~5.0 | 2.0~8.0 | 3.0~12.0 |
Balloon Length | mm | 10~330 | 10~330 | 10~330 |
Coating | Hydrophilic coating, Customized |
The PKP balloon catheter mainly consists of a balloon, developing ring, catheter (consisting of an outer tube and an inner tube), support wire, Y-fitting, and check valve (if applicable).
High voltage withstand
Excellent puncture resistance
● PKP balloon catheter is used as an auxiliary instrument for vertebroplasty and kyphoplasty to restore vertebral height.
Unit | Typical Value | |
Balloon Diameter | mm | 6~17, Customized |
Balloon Length | mm | 8~22, Customized |
Filling Pressure | psi | ≥700 |
Working Channel Dimensions | mm | 3.0, 3.5 |
Rated Burst Pressure (RBP) | atm | ≥11 |